The vmdet recipe¶
- vmdet¶
Synopsis¶
Determine RON, gain, and bad pixels from a set of flat field exposures.
Description¶
This recipe is used to estimate the read-out-noise (RON) and the gain of the CCD, and to determine the positions of the bad pixels.
The input SOF should contain at least five pairs of flat field exposures, all belonging to the same quadrant, each pair corresponding to a different exposure time. The flat fields can be produced either in imaging or in MOS mode. In MOS mode a HR grism is used, in order to illuminate the CCD also beyond the central region used in direct imaging mode, but no mask is inserted at the telescope focal plane.
This type of exposure cannot really be considered a spectral flat field, because the CCD is exposed to white light (i.e., without a wavelength dependency along the dispersion direction). The flat fields generated for the purpose of determining the detector properties are assigned the DO category DETECTOR_PROPERTIES, to distinguish them from the more common IMG_SCREEN_FLAT or MOS_SCREEN_FLAT that are used to produce master calibrations.
Input files¶
DO category: Type: Explanation: Required:
DETECTOR_PROPERTIES Raw Flat field Y
MASTER_BIAS Calib Master bias Y
Output files¶
DO category: Data type: Explanation:
CCD_TABLE FITS table Bad pixel table
(none) FITS image Bad pixel image
(none) FITS image Error image
Only the primary product, the bad pixel table, is copied (or moved) to the product directory. Other products are generated only on request (typically for debug purposes) and are not assigned a DO category as they would not be used anywhere in further data processing steps.
For more details, please refer to the VIMOS Pipeline User’s Guide.
Constructor¶
- cpl.Recipe("vmdet")
Create an object for the recipe vmdet.
import cpl
vmdet = cpl.Recipe("vmdet")
Parameters¶
- vmdet.param.CreateBadPixelMap¶
Bad pixel map creation (bool; default: True) [default=True].
- vmdet.param.CreateErrorImage¶
Response slope uncertainty map creation (bool; default: False) [default=False].
- vmdet.param.DetectionThreshold¶
Threshold in determination of bad pixels (float; default: 5.0) [default=5.0].
- vmdet.param.DetectionMode¶
Bad pixel detection method (str; default: ‘Intolerant’) [default=”Intolerant”].
- vmdet.param.ComputeQC¶
Compute QC1 parameters. (bool; default: True) [default=True].
The following code snippet shows the default settings for the available parameters.
import cpl
vmdet = cpl.Recipe("vmdet")
vmdet.param.CreateBadPixelMap = True
vmdet.param.CreateErrorImage = False
vmdet.param.DetectionThreshold = 5.0
vmdet.param.DetectionMode = "Intolerant"
vmdet.param.ComputeQC = True
You may also set or overwrite some or all parameters by the recipe parameter param, as shown in the following example:
import cpl
vmdet = cpl.Recipe("vmdet")
[...]
res = vmdet( ..., param = {"CreateBadPixelMap":True, "CreateErrorImage":False})
See also
cpl.Recipe for more information about the recipe object.
Bug reports¶
Please report any problems to ESO VIMOS Pipeline Team. Alternatively, you may send a report to the ESO User Support Department.
Copyright¶
This file is part of the VIMOS Instrument Pipeline Copyright (C) 2002-2005 European Southern Observatory
This program is free software; you can redistribute it and/or modify it under the terms of the GNU General Public License as published by the Free Software Foundation; either version 2 of the License, or (at your option) any later version.
This program is distributed in the hope that it will be useful, but WITHOUT ANY WARRANTY; without even the implied warranty of MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the GNU General Public License for more details.
You should have received a copy of the GNU General Public License along with this program; if not, write to the Free Software Foundation, Inc., 51 Franklin St, Fifth Floor, Boston, MA 02110-1301 USA
Code author: ESO VIMOS Pipeline Team <https://support.eso.org>